| Room Size | |
| Minimum | 6 x 7 feet |
| Typical | 8 x 10 feet |
| Camera Footprint | 100 x 120 cm / 47 x 39 in |
| Total Camera Weight | Approx. 444 kg |
| Floor Loading | 0.7 lbs per square inch (without patient) |
Patient Imaging Chair
| Maximum Patient Weight | 130 kg / 350 lbs |
| Vertical Travel | 14 cm / 5.5 in |
Power Requirements
| 120/220V @ 15A dedicated circuit breaker | |
| Power consumption | |
| Maximum | 375 watts |
| Typical | Approx. 150 watts (includes computer) |
Environmental
| Operating Temperature | |
| Range | 60°F to 80°F (16°C to 27°C) |
| Maximum variation | 10°F / Hour (5.5°C / Hour) |
| Humidity | 30% - 80%, non-condensing |
| Max heat dissipation | 2,500 BTU / hour |