The CardiArc SPECT System
General specifications
Physical
Room Size
Minimum 6 x 7 feet
Typical 8 x 10 feet
Camera Footprint 100 x 120 cm / 47 x 39 in
Total Camera Weight Approx. 444 kg
Floor Loading 0.7 lbs per square inch (without patient)


Patient Imaging Chair
Maximum Patient Weight 130 kg / 350 lbs
Vertical Travel 14 cm / 5.5 in


Power Requirements
120/220V @ 15A dedicated circuit breaker
Power consumption
Maximum 375 watts
Typical Approx. 150 watts (includes computer)


Environmental
Operating Temperature
Range 60°F to 80°F (16°C to 27°C)
Maximum variation 10°F / Hour (5.5°C / Hour)
Humidity 30% - 80%, non-condensing
Max heat dissipation 2,500 BTU / hour